开元|中国有限公司官网

Hotline

Move the X-ray machine

一、 Mobile X-ray machine features: 1.For hospital and emergency use 2.Flexible mobile operating performance 3.Wireless remote exposure, greatly reducing the dose of physician radiation 二. The main technical parameters 1.Power supply……
Product Introduction

一、Mobile X-ray machine features:

1.For hospital and emergency use

2.Flexible mobile operating performance

3.Wireless remote exposure, greatly reducing the dose of physician radiation

二.The main technical parameters

1.Power supply conditions and operating mode

voltage: communicate with 220土22V

Power frequency:  50Hz±0.5Hz;

battery capacity:≥8KVA;

Power supply allows maximum internal resistance:1Ω  

Operation mode:Intermittent loading continuous operation

2、the highest rated capacity see Table 1

Tube current(mA)  Tube voltage      time

   15                    90         6.3

   30                    90         6.3

   60                    90         4.0

   100                   80         3.2

3、Photography conditions:   Tube voltage:50-90kV

                                 Tube current:15、30、60、100mA

                                 time:0.08s-6.3 s a total of 19 files, according to R10 'coefficient selected.

4、Maximum output power:

   80kV  100mA  0.1S  5.92KVA.

5、Nominal electric power

   90kV  60mA  0.1S  4.00KVA.

6、input power:5.92KVA.

7、Mechanical behavior

      X-ray generator focus to the ground up and down activities range of 810-1790mm

      X-ray generator transverse stretching range of 255mm

      The left and right swing angles are ± 50º  

The X-ray generator has a rotation angle of ± 180º around the horizontal axis  

The X-ray generator has a rotation angle of ± 90º around the vertical axis;

Move the X ray machine

prevMobile X-ray machine
nextDigital moving X-ray machine